Palwai A, Skowronski M, Coreno A, Drummond C, McFadden ER

Critical comparisons of the clinical performance of oxygen-conserving devices.
(Am J Respir Crit Care Med)
RATIONALE: Clinical testing of oxygen-conserving devices is not mandated before marketing. Consequently, little is known about individual or comparative therapeutic effectiveness. OBJECTIVES: To relate oxygen delivery from prototypical instruments to physiological performance. METHODS: Thirteen subjects with obstructive lung disease performed progressive treadmill exercise while inhaling either room air, 2 L O(2)/min, or bolus oxygen from four commercially available conserving devices at regulator settings of 2, 5, and continuous. The devices were studied blindly in random order after first being tested to determine performance characteristics. Pulse oximetry, oxygen delivery, and nasal and oral ventilations were monitored at rest and with exertion. MEASUREMENTS AND MAIN

RESULTS: At a setting of 2 at rest, all conservers maintained saturation greater than 90%, but there were significant differences in oxygenation between systems. Only one equaled 2 L O(2)/min. With exertion, saturation decreased with all conservers but not with 2 L O(2)/min. One device did not perform any better than room air. Two systems provided less oxygen than predicted, one more, and in one the expected and actual amounts were equal only at rest. Breath-by-breath performance was highly variable, with irregular activation and inconsistent oxygen bolus size delivery. Increasing oxygen pulse volume to the point of eradicating conservation with the continuous setting did not eliminate all disparities.

CONCLUSIONS: The mechanical and clinical performances of current oxygen conservers are highly variable and in some instances actually contribute to limitations in exercise ability. Seemingly equivalent technical features do not guarantee equivalent therapeutic functionality.

-> Fulltext - The link to the fulltext is for registered users available. Register here for free!
Already registered? -> Login here

Read more articles from the authors:
1. Palwai A
2. Skowronski M
3. Coreno A
4. Drummond C
5. McFadden ER


Related articles

Read abstract in PubMed

Do search in PubMed

Copy citation:


Add to Yahoo    Add to Del.icio.us    Add to Google    Add to AOL    Add to Furl    Add to Connotea    Add to Citeulike    Add to Stumbleupon    Add to Dissect medicine    Add to 2collab   

Add Searching Medicine to your search engines :: Put Searching Medicine on your website :: Make Searching Medicine your homepage :: Add Searching Medicine to your favorites
© 2010 Searching Medicine